Aerosol Instrumentation
Hygroscopic Tandem Differential Mobility Analyser
(Model: TSI HTDMA)
The Hygroscopic Tandem Differential Mobility Analyser (HTDMA) is used to measure size resolved aerosol hygroscopic properties. It can operate in the RH range of 15% to 94% to measure increasing as well as decreasing RH growth factors.
Plasma Torch Aerosol Generator
It is a non- transferred type torch having a power of 40 kW. Argon gas is used as primary plasma gas for plasma generation and once the flame is stabilized, nitrogen gas is added to increase its enthalpy. Water cooling circuit consists of chiller of ~5-7 ton capacity to provide cooling for torch as well as power supply. It not only yields an intense and continuous aerosol stream but can also generate aerosols from a wide variety of materials such as metals, ceramics, and composites.
Powder feeder
(Model: MEC PF-3350)
It is used for feeding metal powders on plasma flame. This powder feeder can be used manually as well as automatically with the Programmable Logic Controller (PLC) system. A motor drive having digital RPM display governs the speed of rotating disk which is proportional to the feed rate of the powder.